Wc. Fowler et al., The free electron laser: a system capable of determining the gold standardin laser vision correction, NUCL INST A, 429(1-3), 1999, pp. 497-501
Introduction. In laser vision correction surgery, lasers are generally util
ized based on their beam-tissue interactions and corneal absorption charact
eristics. Therefore, the free electron laser, with its ability to provide b
road wavelength tunability, is a unique research tool for investigating wav
elengths of possible corneal ablation, Methods. Mark III free electron lase
r wavelengths between 2.94 and 6.7 mu m were delivered in serial 0.1 mu m i
ntervals to corneas of freshly enucleated porcine globes. Collateral damage
, ablation depth, and ablation diameter were measured in histologic section
s. Results. The least collateral damage (12-13 mu m) was demonstrated at th
ree wavelengths: 6.0, 6.1 (amide I), and 6.3 mu m. Minimal collateral damag
e (15 mu m) was noted at 2.94 mu m (OH-stretch) and at 6.2 mu m. Slightly g
reater collateral damage was noted at 6.45 mu m (amide II), as well as at t
he 5.5-5.7 mu m range, but this was still substantially less than the colla
teral damage noted at the other wavelengths tested. Conclusions. Our result
s suggest that select mid-infrared wavelengths have potential for keratoref
ractive surgery and warrant additional study. Further, the free electron la
ser's ability to allow parameter adjustment in the farultraviolet spectrum
may provide unprecedented insights toward establishing the gold-standard pa
rameters for laser vision correction surgery. (C) 1999 Elsevier Science B.V
. All rights reserved.