Tuned-circuit differential susceptometer for high-pressure and low-temperature studies of small organic crystals - A new way of measuring quantum oscillations
Ms. Nam et al., Tuned-circuit differential susceptometer for high-pressure and low-temperature studies of small organic crystals - A new way of measuring quantum oscillations, SYNTH METAL, 103(1-3), 1999, pp. 2259-2260
A new technique has been developed to measure the differential susceptibili
ty of small samples under pressure using a tuned circuit. The very low heat
dissipation in the experiment means that the technique can be transferred
easily to a dilution fridge. Our setup provides a simple and versatile way
of measuring the de Haas-van Alphen (dHvA) effect under pressure.