S. Karthikeyan et al., Determination of arsenic in sea water by sorbent extraction with hydride generation atomic absorption spectrometry, TALANTA, 49(3), 1999, pp. 523-530
A rapid and sensitive sorbent extraction hydride generation-flow injection
analysis atomic absorption spectrometric (HG-FIAS-AAS) method is described
for the determination of As(III) and As(V) based upon online preconcentrati
on on a microcolumn packed with activated alumina. In the present procedure
these arsenicals are complexed with quinolin-8-ol-5-sulphonic acid from ne
utral solutions in the flow injection system and adsorbed on the column. Th
e preconcentrated species are eluted with 10% HCl, mixed with 0.5% sodium b
orohydride and carried to the HG-FIAS cell with a carrier gas flow rate of
75 ml min(-1). The retention efficiency is found to be better than 98% with
sensitivity enhancement of 12 and 10 for As(III) and As(V), respectively,
for a 20 s preconcentration period. The respective detection limits are 0.0
5 and 2 ng ml(-1) for As(III) and As(V). The throughput of the samples is f
ound to be 60 h(-1), with a loading time of 20 s. The method has been appli
ed to sea water samples. (C) 1999 Elsevier Science B.V. All rights reserved
.