Localized visualization of chemical cross-talk in microsensor arrays by using scanning electrochemical microscopy

Citation
Dj. Strike et al., Localized visualization of chemical cross-talk in microsensor arrays by using scanning electrochemical microscopy, MIKROCH ACT, 131(1-2), 1999, pp. 47-55
Citations number
25
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
MIKROCHIMICA ACTA
ISSN journal
00263672 → ACNP
Volume
131
Issue
1-2
Year of publication
1999
Pages
47 - 55
Database
ISI
SICI code
0026-3672(1999)131:1-2<47:LVOCCI>2.0.ZU;2-X
Abstract
An investigation of an array of four Pt microband electrodes 25 mu m wide a nd spaced 25 mu m apart was performed with the scanning electrochemical mic roscope (SECM). Where possible the SECM measurements were confirmed with co nventional electrochemical measurements. It is shown how the sensitivity of the SECM recycling current to the activity of the underlying surface can b e used to probe the homogeneity of enzyme-modified microelectrodes. The dif fusion of H2O2 between these micro enzyme-electrodes and unmodified electro des was investigated and it was demonstrated how the SECM can be a powerful tool in the elucidation of the properties of these electrodes.