Background Implantable cardioverter-defibrillation function is critically d
ependent on reliable sensing of intracordiac signals. Lead systems that use
integrated sensing, in which the distal shocking coil is part of both the
sensing and shocking pathways, may be prone to undersensing of ventricular
fibrillation, especially during redetection after a failed first shock. To
assess the effect of endocardial lead system on redetection, we compared a
dedicated rate-sensing lead and 2 generations of integrated sensing defibri
llator leads with a uniform testing algorithm and pulse generator.
Methods The study group consisted of 72 patients when implantable cardiover
ier-defibrillator implatation. Three transvenous rate-sensing leads were ev
aluated: a standard pacing lead, incorporating true bipolar sensing without
ventricular coils, or an integrated shocking and sensing lead (Endotak C)
with either 6-mm (60 series) or 12-mm (70 series) spacing between the sensi
ng tip and shocking coil. Redetection was assessed from a failed first shoc
k lust below defibrillation threshold.
Results Compared with the dedicated bipolar lead, redetection was prolonged
with the 60 series lead (8.3 +/- 3.6 vs 6.6 +/- 2.3 seconds, P = .04). Mor
eover, prolonged redetection (>8 seconds) was observed in 41% of patients w
ith 60 series leads compared with only 11% with dedicated bipolar leads (P
< .01). No significant effects on redetection were noted with an integrated
lead with greater spacing between the tip and coil (70 series).
Conclusions Delayed redetection is frequently noted with an integrated lead
with close spacing between the tip and coil. Detailed evaluation of detect
ion and redetection of these leads should be performed at the time of pulse
generator replacement.