A 3-D Monte Carlo simulation of a small animal Positron Emission Tomographwith millimeter spatial resolution

Citation
A. Bevilacqua et al., A 3-D Monte Carlo simulation of a small animal Positron Emission Tomographwith millimeter spatial resolution, IEEE NUCL S, 46(3), 1999, pp. 697-701
Citations number
11
Categorie Soggetti
Apllied Physucs/Condensed Matter/Materiales Science","Nuclear Emgineering
Journal title
IEEE TRANSACTIONS ON NUCLEAR SCIENCE
ISSN journal
00189499 → ACNP
Volume
46
Issue
3
Year of publication
1999
Part
2
Pages
697 - 701
Database
ISI
SICI code
0018-9499(199906)46:3<697:A3MCSO>2.0.ZU;2-N
Abstract
We have used the EGS4 Monte Carlo code to simulate the performance of a Pos itron Emission Tomograph for small animals. The tomograph is a four-head de tector, made of matrices of 400 YAP:Ce finger-like crystals (2x2x30 mm(3) e ach) coupled to Position Sensitive Photo-Multipliers Hamamatsu R2486-06. 3- D reconstruction algorithms have been applied to the 3-D simulated data. Th e reconstructed volume was 2x2x2 cm(3), the same of the reference mouse bra in. The Monte Carlo results show that a spatial resolution below 2 mm (EWHM ) can be obtained over the 2x2x2 cm3 volume with a sensitivity of about 600 counts/(mu Ci.s) for a gantry diameter of 15 cm, and a threshold of 50 keV .