Pulse height distributions of beta rays in C-14 measurement with liquid scintillation counting system quantulus

Citation
A. Suzuki et al., Pulse height distributions of beta rays in C-14 measurement with liquid scintillation counting system quantulus, IEEE NUCL S, 46(3), 1999, pp. 302-305
Citations number
10
Categorie Soggetti
Apllied Physucs/Condensed Matter/Materiales Science","Nuclear Emgineering
Journal title
IEEE TRANSACTIONS ON NUCLEAR SCIENCE
ISSN journal
00189499 → ACNP
Volume
46
Issue
3
Year of publication
1999
Part
1
Pages
302 - 305
Database
ISI
SICI code
0018-9499(199906)46:3<302:PHDOBR>2.0.ZU;2-J
Abstract
To investigate the variations of the solar activity over a long time scale, we have started the measurements of C-14 in an ancient tree, synthesizing 10 g benzene from the tree rings and using the,liquid scintillation countin g system Quantulus with ultra low background level. The background rate of the Quantulus was 1.2 cpm while the counting rate of C-14 was 91.0 cpm for the ancient tree The pulse height distribution of beta rays from C-14 was w ell fitted with theoretical beta rays spectrum. The Quantulus detected beta rays down to 7.5 keV with good detection efficiency. It was confirmed that the total counts from the C-14 in the sample benzene can be measured withi n 0.2% accuracy on the basis of the pulse height distribution of a standard benzene sample, using the least squares fitting method.