A concentric-ring equivalent phased array method to model fields of large axisymmetric ultrasound transducers

Citation
Xb. Fan et al., A concentric-ring equivalent phased array method to model fields of large axisymmetric ultrasound transducers, IEEE ULTRAS, 46(4), 1999, pp. 830-841
Citations number
35
Categorie Soggetti
Optics & Acoustics
Journal title
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL
ISSN journal
08853010 → ACNP
Volume
46
Issue
4
Year of publication
1999
Pages
830 - 841
Database
ISI
SICI code
0885-3010(199907)46:4<830:ACEPAM>2.0.ZU;2-4
Abstract
An equivalent concentric-ring ultrasound phased array method was developed to estimate ultrasonic continuous wave fields generated by axisymmetric sin gle-source transducers, The method models a given source as a concentric-ri ng phased array by mathematically segmenting it into many rings and subsequ ently finding the amplitude and phase for each ring that produces an acoust ic field similar to the held of the single-source transducer. The excitatio n source of each ring was calculated using an inverse technique based on co mplex pressure measurements along a radial line close to the source. The pr edicting abilities of the method are evaluated by comparing measured and es timated ultrasound fields for six different transducers. The results show t hat the concentric-ring equivalent phased array method (CREPAM) is able to estimate quantitatively the ultrasound fields generated by large axisymmetr ic planar and focused transducers.