On-line trace enrichment for the simultaneous determination of microcystins in aqueous samples using high-performance liquid chromatography with diode-array detection
Hs. Lee et al., On-line trace enrichment for the simultaneous determination of microcystins in aqueous samples using high-performance liquid chromatography with diode-array detection, J CHROMAT A, 848(1-2), 1999, pp. 179-184
The need for a rapid, sensitive and reliable analytical method for cyanobac
terial toxins, microcystins, has been emphasized by the awareness of toxic
cyanobacteria as a human-health risk through drinking water. A new high-per
formance liquid chromatographic method with column switching was developed
for the determination of microcystin-LR, -RR and -YR from water samples wit
hout pre-purification. The filtered water sample was passed through a Zorba
x CN precolumn at a flow-rate of 3 ml/min for on-line trace enrichment. Aft
er valve switching, concentrated analytes were eluted in back-flush mode an
d separated on a Luna C-18 column with a gradient of acetonitrile -20 mM ph
osphate buffer (pH 2.5). The method showed excellent precision, accuracy an
d speed with detection limits of 0.02 mu g/ml from 100 ml of surface water.
The total analysis time per sample was about 90 min. This method improves
reliability, sensitivity and sample throughput, and shortens the analysis t
ime compared to analysis methods using off-line solid-phase extraction. (C)
1999 Elsevier Science B.V. All rights reserved.