Aa. Nava-ocampo et al., Simplified method to quantify furosemide in urine by high-performance liquid chromatography and ultraviolet detection, J CHROMAT B, 730(1), 1999, pp. 49-54
Simplified reversed-phase high-performance liquid chromatographic method wi
th ultraviolet detection at 280 nm without extraction procedure is describe
d to quantify furosemide in rabbit and human urine. An internal standard wa
s not used. The lower limit of quantitation was 0.750 mu g/ml using 50 mu l
urine samples (100 mu l of total injection volume), and Linear response wa
s tested from 0.750 to 250 mu g/ml in both humans and rabbits. Within and b
etween-day accuracy and precision were always below 10% at all analyzed con
centrations. Validation data showed that this method is linear, sensitive,
selective, specific, accurate and reproducible. (C) 1999 Published by Elsev
ier Science B.V. All rights reserved.