Quantitative imaging of perfusion using a single subtraction, second versio
n (QUIPSS II) is a pulsed arterial spin labeling (ASL) technique for improv
ing the quantitation of perfusion imaging by minimizing two major systemati
c errors: the variable transit delay from the distal edge of the tagged reg
ion to the imaging slices, and the contamination by intravascular signal fr
om tagged blood that flows through the imaging slices. However, residual er
rors remain due to incomplete saturation of spins over the slab-shaped tagg
ed region by the QUIPSS II saturation pulse, and spatial mismatch of the di
stal edge of the saturation and inversion slice profiles. By replacing the
original QUIPSS II saturation pulse with a train of thin-slice periodic sat
uration pulses applied at the distal end of the tagged region, the accuracy
of perfusion quantitation is improved. Results of single and multislice st
udies are reported, (C) 1999 Wiley-Liss, Inc.