J. Felblinger et al., Restoration of electrophysiological signals distorted by inductive effectsof magnetic field gradients during MR sequences, MAGN RES M, 41(4), 1999, pp. 715-721
Citations number
19
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
A generally applicable method for almost complete suppression of signal art
ifacts on electrophysiological signals caused by B-0-gradient switching (gr
adient noise) is presented. The method is demonstrated for electrocardiogra
ms (ECGs) but can also be used for other electrophysiological signals. It t
akes advantage of the fact that under certain conditions, the effect of swi
tching the Be-field gradient upon an electrophysiological signal can be mod
eled as a linear time-invariant system and fully characterized by pulse res
ponse functions. It is shown how the system's pulse response functions of t
he X, Y, and Z gradients can be determined and how gradient noise can be el
iminated efficiently. The elimination of gradient noise by the proposed met
hod causes in the current arrangement a constant delay of 128 msec, which i
s acceptable for patient monitoring and magnetic resonance sequence trigger
ing. (C) 1999 Wiley-Liss, Inc.