K. Stierstorfer et M. Spahn, Self-normalizing method to measure the detective quantum efficiency of a wide range of x-ray detectors, MED PHYS, 26(7), 1999, pp. 1312-1319
Citations number
19
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
The detective quantum efficiency (DQE) is widely accepted as the most relev
ant parameter to characterize the image quality of medical x-ray systems. I
n this article we describe a solid. method to measure the DQE. The strength
of the method lies in the fact that it is self-normalizing so measurements
at very low spatial frequencies are not needed. Furthermore, it works on a
ny system with a response function which is linear in the small-signal appr
oximation. We decompose the DQE into several easily accessible quantities a
nd discuss in detail how they can be measured. At the end we lead the inter
ested reader through an example. Noise equivalent quanta and normalized con
trast values are tabulated for standard radiation qualities. (C) 1999 Ameri
can Association of Physicists in Medicine. [S0094-2405(99)00907-4].