T. Seppala et al., Modelling of brain tissue substitutes for phantom materials in neutron capture therapy (NCT) dosimetry, RADIAT PH C, 55(3), 1999, pp. 239-246
The aim of this study was to define the most suitable brain tissue phantom
material for neutron capture therapy (NCT), The calculated distributions of
the thermal neutron fluence and the fast neutron and gamma dose in normal
brain tissue defined by ICRU were compared to those in four brain tissue su
bstitutes: water; PMMA and two normal brain tissue substitute liquids, A an
d B, defined by us. Liquid B is an excellent material to simulate the neutr
on and gamma interactions in normal brain tissue and it can be considered f
or use in situations for which higher accuracy is required, although it may
not be suitable for routine use. Water is a suitable brain tissue substitu
te for periodic epithermal beam calibration, for routine quality control an
d for intercomparison of beams in NCT dosimetry. (C) 1999 Elsevier Science
Ltd. All rights reserved.