Mp. Aguilar-caballos et al., Simultaneous determination of benzoic acid and saccharin in soft drinks byusing lanthanide-sensitized luminescence, ANALYST, 124(7), 1999, pp. 1079-1084
A simple and fast approach is used for the first time to develop a time res
olved lanthanide-sensitized luminescence method for the simultaneous determ
ination of a preservative and a sweetener, namely benzoic acid (BZ) and sac
charin (SC), respectively, in food samples. The method involves the formati
on of the corresponding ternary chelates with terbium(III) and trioctylphos
phine oxide (TOPO) in the presence of Triton X-100, and the measurement of
the initial rate and equilibrium signal of this system, which were obtained
in 0.1 and 5 s, respectively. The dynamic ranges of the calibration graphs
, obtained by using kinetic and equilibrium measurements, were 0.2-36 mu g
ml(-1) and 0.15-30 mu g ml(-1), respectively, for BZ, and 3.3-24 mu g ml(-1
) and 4-36 mu g ml(-1) for SC and the detection limits were 0.07 and 0.04 m
u g ml(-1), respectively, for BZ, and 1.1 and 1.2 mu g ml(-1), respectively
, for sodium SC. The relative standard deviation ranged between 2.3 and 3.0
%; Both compounds were determined simultaneously by using a system of two e
quations which were resolved by using the calibration data obtained individ
ually for each analyte and by multiple linear regression. Mixtures of BZ an
d SC in ratios between 3:1 and 1:9 were satisfactorily resolved by using bo
th approaches. The method was applied to the direct analysis of several sof
t drinks. Analytical recoveries ranged between 89.3 and 108.5%.