B. Nowack et U. Von Gunten, Determination of chlorate at low mu g/1 levels by ion-chromatography with postcolumn reaction, J CHROMAT A, 849(1), 1999, pp. 209-215
A new method for the determination of low concentrations of chlorate in nat
ural waters is described. Chlorate is analyzed by ion-chromatography follow
ed by an osmate-catalyzed postcolumn reaction of chlorate with iodide and U
V-detection of triiodide. The new osmate catalysis allows to carry out the
oxidation of iodide by chlorate at pH 3 instead of 6 M HCl for the uncataly
zed reaction. A detection limit of 5 nM (0.4 mu g/l) chlorate is achieved.
The method also allows the simultaneous determination of chlorite, bromate,
and nitrite at the low mu g/l level. (C) 1999 Elsevier Science B.V. All ri
ghts reserved.