National comparison of hydrometer calibrations

Citation
H. Adametz et al., National comparison of hydrometer calibrations, PTB-MITT, 109(3), 1999, pp. 138-145
Citations number
8
Categorie Soggetti
Instrumentation & Measurement
Journal title
PTB-MITTEILUNGEN
ISSN journal
0030834X → ACNP
Volume
109
Issue
3
Year of publication
1999
Pages
138 - 145
Database
ISI
SICI code
0030-834X(199906)109:3<138:NCOHC>2.0.ZU;2-Y
Abstract
A national comparison of hydrometer calibrations was performed by the verif ication offices in Berlin, Ilmenau, Munchen and Wertheim and by the Physika lisch-Technische bundesanstalt (PTB) which acted as reference laboratory. F our hydrometers for density and concentration with a high resolution and in tended for use in different liquids were calibrated at three scale lines ea ch. Correction values for two hydrometers were given for different referenc e temperatures (15 degrees C and 20 degrees C) and for reading both in the liquid plane and at the upper level of the liquid meniscus. Two verificatio n offices and the PTB used the Cuckow method. The other laboratories compar ed the hydrometers in appropriate liquids with standard hydrometers. The 10 7 correction values determined by the verification offices agree with the P TB's reference values within the stated uncertainties The constancy of hydr ometers is demonstrated by the correction values of one hydrometer over a p eriod of 26 years.