A FIA procedure for enzymatic determination of high glucose concentrations. Application to parenteral solutions analysis

Citation
E. Barrado et al., A FIA procedure for enzymatic determination of high glucose concentrations. Application to parenteral solutions analysis, ANALUSIS, 27(2), 1999, pp. 170-174
Citations number
34
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
ANALUSIS
ISSN journal
03654877 → ACNP
Volume
27
Issue
2
Year of publication
1999
Pages
170 - 174
Database
ISI
SICI code
0365-4877(199903)27:2<170:AFPFED>2.0.ZU;2-W
Abstract
A flow injection system for the determination of high levels of glucose in parenteral solutions based on amperometric detection of hydrogen peroxide p roduced in the glucose oxidase reaction was developed, The determinations w ere made without any sample pre-treatment. The flow injection manifold incl udes two dialysis units to perform dilution of the samples and to maintain the enzyme in a closed re-circulation loop. Due to the hydrodynamic charact eristics of the analytical system, the variables were optimized using a par ameter design of Taguchi. Under optimal conditions, a linear response for g lucose concentrations, between 0.1 and 1.0 M was achieved. A good reproduci bility (r.s.d. < 3%) and a sampling rate of 30 determinations/hour were obs erved. The results obtained are in a good agreement with those obtained usi ng the spectrophotometric enzymatic method.