Vc. Dall'Orto et al., Comparison between three amperometric sensors for phenol determination in olive oil samples, ANAL LETTER, 32(10), 1999, pp. 1981-1990
The analytical performance of two modified electrodes (a tyrosinase biosens
or and a glassy carbon electrode (GCE) covered with poly[Ni(protoporphyrin
IX)dimethyl ester]) in comparison with the bare electrode has been evaluate
d in a flow injection system using standard solutions of oleuropein and oli
ve oil samples. The tyrosinase biosensor appeared to be an appropriate devi
ce for the selective determination of phenols, moreover it can be easily in
corporated in an on-line system. This method was used to evaluate the stabi
lity of the samples.