X. Yu et al., Study of the simultaneous determination of trace arsenic and mercury by flow injection/hydride generation/ICP-AES, ANAL LETTER, 32(10), 1999, pp. 2105-2114
A combined system of flow injection/hydride generation coupled with multich
annel plasma spectrometer was put forward and applied to the simultaneous d
etermination of trace arsenic and mercury. The interface between flow injec
tion hydride generation (FI-HG) and ICP-AES, gas/liquid separator and main
experimental parameters were described. The detection limits of the propose
d method could be achieved to 0.x mu g/L level, The RSD 1.4% and 1.3% at 10
ng/mL levels of As and Hg respectively. The frequency of sampling was 150/
h with 80 mu L sampling volume. The proposed method has been applied to the
simultaneous determination of As and Hg in some real samples.