SIMULTANEOUS QUANTIFICATION OF INSECTICIDES INCLUDING CARBARYL IN DRINKING-WATER BY GAS-CHROMATOGRAPHY USING DUAL ELECTRON-CAPTURE AND NITROGEN-PHOSPHORUS DETECTION
Ys. Ohshin et al., SIMULTANEOUS QUANTIFICATION OF INSECTICIDES INCLUDING CARBARYL IN DRINKING-WATER BY GAS-CHROMATOGRAPHY USING DUAL ELECTRON-CAPTURE AND NITROGEN-PHOSPHORUS DETECTION, Journal of chromatography, 769(2), 1997, pp. 285-291
Citations number
15
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
A rapid and simple gas chromatographic method for the simultaneous det
ermination of malathion, parathion, fenitrothion, diazinon and carbary
l in drinking water is presented. A fused-silica SE-54 column was used
for the separation of the insecticides and was split into two specifi
c detectors; electron-capture and nitrogen-phosphorus detectors by usi
ng column switching. A water sample was extracted with methylene chlor
ide. The organic phase was evaporated and the residue was derivatized
with pentafluoropropionic acid anhydride. The detection limit was belo
w 0.1 ng/ml and the calibration curves showed good linearity with r=0.
998 similar to 1.000. The method was sensitive, reproducible and simpl
e enough to permit the reliable routine analysis of the pesticides in
drinking water.