The successful application of single-lead VDD pacing during the last few ye
ars hers generated the idea of single-lead DDD pacing: Preliminary data fro
m several single-lead VDD studies attempting to pace the atrium by a floati
ng atrial dipole are unsatisfactory, causing an unacceptably high current d
rain of the device. We studied thefeasibility as well as the short- and lon
g-term stability of atrioventricular sequential pacing, using a new single-
pass, tined DDD lead. In eight consecutive patients (age 73 +/- 16 years) w
ith symptomatic higher degree A V block and intact sinus node function, thi
s new single-pass DDD lead was implanted in combination with a. DDDR pacema
ker. Correct VDD and DDD function wets studied at implantation; at discharg
e; and at 1, 3, and 6 months of follow-up. At implant, the atrial stimulati
on threshold was 0.6 +/- 0.2 V/0.5 ms. During follow-up, the atrial pacing
thresholds in different everyday positions averaged 2.1 +/- 0.5 V at discha
rge, 2.9 +/- 0.5 V at 2 month, 3.8 +/- 0.4 V at 3 months, and 3.4 +/- 0.4 V
at 6 months (pulse width always 0.5 ms). The measured P wave amplitude at
implantation was 4.5 +/- 2.2 mV; during follow-up the telemetered atrial se
nsitivity thresholds averaged 2.2 +/- 0.3 mV. Phrenic nerve stimulation at
high output pacing (5.0 V/0.5 ms) was observed in three (38%) patients at d
ischarge and in one (13%) patient during follow-up; an intermittent unmeasu
rable atrial lead impedance at 3 and 6 months follow-up was documented in o
ne (13%) patient. This study confirms the possibility of short- and long-te
rm DDD pacing using a single-pass DDD lead. Since atrial stimulation thresh
olds are still relatively high compared to conventional dual-lead DDD pacin
g, further improvements of the atrial electrodes are desirable, enabling lo
wer pacing thresholds and optimizing energy requirements as well as minimiz
ing the potential disadvantage of phrenic nerve stimulation.