Calibration of strontium-90 eye applicator using a strontium external beamstandard

Citation
D. Siddle et K. Langmack, Calibration of strontium-90 eye applicator using a strontium external beamstandard, PHYS MED BI, 44(7), 1999, pp. 1597-1608
Citations number
24
Categorie Soggetti
Multidisciplinary
Journal title
PHYSICS IN MEDICINE AND BIOLOGY
ISSN journal
00319155 → ACNP
Volume
44
Issue
7
Year of publication
1999
Pages
1597 - 1608
Database
ISI
SICI code
0031-9155(199907)44:7<1597:COSEAU>2.0.ZU;2-3
Abstract
Four techniques for measuring the dose rate from Sr-90 concave eye plaques are presented. The techniques involve calibrating a concave eye plaque agai nst a Sr-90 teletherapy unit using X-Omat him, radiochromic film, black LiF TLD discs and LiF chips. The mean dose rate predicted by these dosimeters is 7.5 cGy s(-1). The dose rate quoted by the manufacturer is 33% lower tha n this value, which is consistent with discrepancies reported by other auth ors. Calibration against a 6 MV linear accelerator was also carried out usi ng each of the above dosimetric devices, and appropriate sensitivity correc tion factors have been presented.