Four techniques for measuring the dose rate from Sr-90 concave eye plaques
are presented. The techniques involve calibrating a concave eye plaque agai
nst a Sr-90 teletherapy unit using X-Omat him, radiochromic film, black LiF
TLD discs and LiF chips. The mean dose rate predicted by these dosimeters
is 7.5 cGy s(-1). The dose rate quoted by the manufacturer is 33% lower tha
n this value, which is consistent with discrepancies reported by other auth
ors. Calibration against a 6 MV linear accelerator was also carried out usi
ng each of the above dosimetric devices, and appropriate sensitivity correc
tion factors have been presented.