The relationship between evoked potentials and measurements of S-100 protein in cerebrospinal fluid during and after thoracoabdominal aortic aneurysmsurgery
Ep. Van Dongen et al., The relationship between evoked potentials and measurements of S-100 protein in cerebrospinal fluid during and after thoracoabdominal aortic aneurysmsurgery, J VASC SURG, 30(2), 1999, pp. 293-300
Citations number
18
Categorie Soggetti
Cardiovascular & Respiratory Systems","Cardiovascular & Hematology Research
Objective: This study was performed to correlate the changes in concentrati
on of S-100 protein in the cerebrospinal fluid (CSF) during and after thora
coabdominal aortic aneurysm (TAAA) surgery with the results of somatosensor
y and motor evoked potential monitoring.
Methods: The study was designed as a prospective study at St Antonius Hospi
tal in Nieuwegein, The Netherlands. The participants were 19 patients who w
ere undergoing elective TAAA surgery. CSP samples for analysis of S-100 pro
tein were drawn after the induction of anesthesia, during the cross-clamp p
eriod of the critical aortic segment, after 5 minutes of reperfusion of thi
s segment, during the closure of the skin, and 24 hours after the closure o
f the skin. In all the patients, continuous intraoperative recording of myo
genic motor potentials evoked by transcranial electrical stimulation (tcMEP
) and somatosensory potentials evoked by stimulation of the postetior tibia
l nerve took place to monitor the integrity of the spinal cord. The operati
ve technique consisted of staged or sequential clamping to maximize the ben
eficial effect of the distal perfusion by the left heart bypass, continuous
CSP drainage to keep the CSF pressure below 10 mm Hg, and moderate hypothe
rmia (32 degrees C rectal temperature). We correlated the measured concentr
ations of S-100 protein in CSF with the results of evoked potential monitor
ing during surgery and the number of intercostals reimplanted and oversewn.
Results: In all the patients, the concentration of S-100 protein was increa
sed in CSF The highest concentration of S-100 protein was found in the CSF
sample taken 5 minutes after reperfusion of the critical aortic segment. Th
ere was a good (negative) correlation between the changes in 5-100 protein
in CSF and the changes in motor evoked potential monitoring during the cros
s-damp period. The best (negative) correlation was detected between the S-1
00 protein elevation in the CSP sample drawn 5 minutes after reperfusion an
d the tcMEP amplitude reduction during clamping (r = -0.73; P = .007). No r
elation was found between the S-100 protein dynamics in CSF and somatosenso
ry evoked potential monitoring. A positive (r = 0.58; P = .05) correlation
was found between the change in tcMEP amplitude during clamping and the num
ber of reattached intercostals. A moderate to good (r = -0.5 to -0.7; P < .
05) correlation between the number of reattached intecostals and the change
s in 5-100 protein concentration in CSF during TAAA surgery was found. Our
data show that transient elevations in S-100 protein after cross clamping a
re larger in those patients with marked decrease in tcMEP from baseline dur
ing the cross-clamp period.
Conclusion: A correlation is shown between an increasing concentration of S
-100 protein in CSF and a reduction in tcMEP amplitude during cross clampin
g of the aorta. The S-100 protein in CSF seems to be a marker of potential
clinical value in the evaluation of the effects of procedures to detect and
reduce spinal cord ischemia.