O. Fatibello et C. Aniceto, Flow injection potentiometric determination of saccharin in dietary products using a tubular ion-selective electrode, LAB ROBOT A, 11(4), 1999, pp. 234-239
The construction and analytical evaluation of a tubular ion-selective elect
rode coated with an ion pair formed between saccharinate anion and toluidin
e blue O cation incorporated on a poly(vinylchloride) (PVC) matrix was cons
tructed and adapted in a flow-injection (FI) system. The optimum experiment
al conditions found were an analytical path of 120 cm, an injection sample
volume of 500 mu L, a pH of 2.5, a flow rate of 2.3 mL min(-1), and a tubul
ar electrode length of 2.5 cm. The linear response to saccharin concentrati
on ranged from 1.0 x 10(-4) to 2.0 x 10(-2) mol L-1, with a slope of - 53.2
+/- 0.4 m V/decade and a detection limit of 8.0 x 10(-5) mol L-1. The anal
ytical frequency was 40 determinations h(-1) , and this system was used wit
h success for saccharin determination in several dietary products. (C) 1999
John Wiley & Sons, Inc.