Flow injection potentiometric determination of saccharin in dietary products using a tubular ion-selective electrode

Citation
O. Fatibello et C. Aniceto, Flow injection potentiometric determination of saccharin in dietary products using a tubular ion-selective electrode, LAB ROBOT A, 11(4), 1999, pp. 234-239
Citations number
8
Categorie Soggetti
AI Robotics and Automatic Control
Journal title
LABORATORY ROBOTICS AND AUTOMATION
ISSN journal
08957533 → ACNP
Volume
11
Issue
4
Year of publication
1999
Pages
234 - 239
Database
ISI
SICI code
0895-7533(1999)11:4<234:FIPDOS>2.0.ZU;2-L
Abstract
The construction and analytical evaluation of a tubular ion-selective elect rode coated with an ion pair formed between saccharinate anion and toluidin e blue O cation incorporated on a poly(vinylchloride) (PVC) matrix was cons tructed and adapted in a flow-injection (FI) system. The optimum experiment al conditions found were an analytical path of 120 cm, an injection sample volume of 500 mu L, a pH of 2.5, a flow rate of 2.3 mL min(-1), and a tubul ar electrode length of 2.5 cm. The linear response to saccharin concentrati on ranged from 1.0 x 10(-4) to 2.0 x 10(-2) mol L-1, with a slope of - 53.2 +/- 0.4 m V/decade and a detection limit of 8.0 x 10(-5) mol L-1. The anal ytical frequency was 40 determinations h(-1) , and this system was used wit h success for saccharin determination in several dietary products. (C) 1999 John Wiley & Sons, Inc.