M. Blaauw, The reference peak areas of the 1995 IAEA test spectra for gamma-ray spectrum analysis programs are absolute and traceable, NUCL INST A, 432(1), 1999, pp. 74-76
A previously validated algorithm for absolute peak area determination was u
sed to verify the reference peak areas supplied with the 1995 IAEA test spe
ctra for gamma-ray spectrometry. These reference peak areas turn out to be
absolute and traceable to a precision of 0.9%. The reference peak areas are
possibly too low by a factor 0.992 +/- 0.009. It is proposed to employ the
test spectra and reference areas to validate the peak areas obtained with
any algorithm in gamma-ray spectrometry. (C) 1999 Elsevier Science B.V. All
rights reserved.