Fine analysis of the calibration curve for a HPGe spectrometer in gamma ray energy measurement

Citation
Zg. Liu et al., Fine analysis of the calibration curve for a HPGe spectrometer in gamma ray energy measurement, NUCL INST A, 432(1), 1999, pp. 122-129
Citations number
6
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
432
Issue
1
Year of publication
1999
Pages
122 - 129
Database
ISI
SICI code
0168-9002(19990801)432:1<122:FAOTCC>2.0.ZU;2-Z
Abstract
A computational method is suggested to analyze gamma ray peak position data of a series of closely shifted spectra of Ir-192 measured by a HPGe spectr ometer and an iteration program, Recovery-98 has been designed to find the fine curve of the spectrometer nonlinearity for energy calibration. The pro gram is applied to check a previous experiment (measuring the gamma energie s of Sr-85, etc.), and the revised result has been reported. (C) 1999 Elsev ier Science B.V. All rights reserved.