T. Akyuz et al., Uranium analysis by neutron induced fissionography method using solid state nuclear track detectors, RADIAT MEAS, 31(1-6), 1999, pp. 245-248
In this study total twenty samples (eight reference materials and twelve se
diment samples) were analysed for their uranium content which is in the ran
ge of 1-17 mu g/g, by neutron induced fissionography (NIF) method using sol
id state nuclear track detectors (SSNTDs) in comparison with the results of
neutron activation analysis (NAA), delayed neutron counting (DNC) techniqu
e or fluorometric method. It is found that NIF method using SSNTDs is very
sensitive for analysis of uranium.