Temperature dependence of the calibration factor of radon and radium determination in water samples by SSNTD

Citation
I. Hunyadi et al., Temperature dependence of the calibration factor of radon and radium determination in water samples by SSNTD, RADIAT MEAS, 31(1-6), 1999, pp. 301-306
Citations number
4
Categorie Soggetti
Apllied Physucs/Condensed Matter/Materiales Science
Journal title
RADIATION MEASUREMENTS
ISSN journal
13504487 → ACNP
Volume
31
Issue
1-6
Year of publication
1999
Pages
301 - 306
Database
ISI
SICI code
1350-4487(199906)31:1-6<301:TDOTCF>2.0.ZU;2-0
Abstract
The sensitivity of a Ra-226 determination method of water samples by SSNTD was measured as a function of storage temperature during exposure. The meth od is based on an etched track type radon monitor, which is closed into a g as permeable foil and is immersed in the water sample. The sample is sealed in a glass vessel and stored for an exposure time of 10-30 days. The sensi tivity increased more than a factor of two when the storage temperature was raised from 2 degrees C to 30 degrees C. Temperature dependence of the par tition coefficient of radon between water and air provides explanation for this dependence. For practical radio- analytical application the temperatur e dependence of the calibration factor is given by fitting the sensitivity data obtained by measuring 226Ra standard solutions (in the activity: conce ntration range of 0.1-48.5 kBq m(-3))at different storage temperatures.