Purpose: Osteosynthesis by means of bioresorbable implants, mostly of self-
reinforced poly-L-lactide (SR-PLLA), has been used in humans for about 10 y
ears. The aim of this study was to examine the controversy between histolog
ical studies confirming fragmentation of the biomaterial and radiological s
tudies showing no breaking of the material.
Material and Methods. Six patients with displaced malleolar fractures opera
tively treated with biodegradable SR-PLLA screws underwent MR examinations
at 1.5 T, immediately postoperatively and after one to two years.
Results: The biodegradable osteosynthetic screws were clearly seen on all M
R images. Of 12 screws, 6 were broken at the final examination (5 syndesmot
ic transfixation screws and 1 screw through the growth cartilage).
Conclusion: The breaking of a biodegradable osteosynthesis is possible to d
ocument on MR images.