CALIBRATION OF A PYROELECTRIC DETECTOR AT 10.6 MU-M WITH THE NATIONAL-INSTITUTE-OF-STANDARDS-AND-TECHNOLOGY HIGH-ACCURACY CRYOGENIC RADIOMETER

Citation
Tr. Gentile et al., CALIBRATION OF A PYROELECTRIC DETECTOR AT 10.6 MU-M WITH THE NATIONAL-INSTITUTE-OF-STANDARDS-AND-TECHNOLOGY HIGH-ACCURACY CRYOGENIC RADIOMETER, Applied optics, 36(16), 1997, pp. 3614-3621
Citations number
13
Categorie Soggetti
Optics
Journal title
ISSN journal
00036935
Volume
36
Issue
16
Year of publication
1997
Pages
3614 - 3621
Database
ISI
SICI code
0003-6935(1997)36:16<3614:COAPDA>2.0.ZU;2-K
Abstract
The National Institute of Standards and Technology (NIST) is establish ing an infrared detector calibration facility to improve radiometric s tandards at infrared wavelengths. The absolute response of the cryogen ic bolometer that serves as the transfer standard for this facility is being linked to the NIST high-accuracy cryogenic radiometer (HACR) at a few laser wavelengths. At the 10.6-mu m CO2 laser line, this Link i s being established through a pyroelectric detector that has been cali brated against the HACR. We describe the apparatus, methods, and uncer tainties for the calibration of this pyroelectric detector.