Steady-state model calculations predicting the influence of key parameterson the lower detection limit and ruggedness of solvent polymeric membrane ion-selective electrodes
T. Zwickl et al., Steady-state model calculations predicting the influence of key parameterson the lower detection limit and ruggedness of solvent polymeric membrane ion-selective electrodes, ELECTROANAL, 11(10-11), 1999, pp. 673-680
Using a previously described steady-state model, the influence of various p
arameters on the response function of ion-selective electrodes is investiga
ted. The parameters studied include activities of interfering ions of the s
ample and inner solution, selectivity coefficients, lipophilicily of the an
alyte, concentrations of ionophore and lipophilic ionic sites in the membra
ne, the thickness of the Nernstian diffusion layer and the membrane as well
as the diffusion coefficients in both phases. The results obtained provide
indications as to how the lower detection limit of potentiometric sensors
and their ruggedness can be improved.