We report 5 cases of severe failure of the polyethylene liner at 4 to 7.4 y
ears after uncemented anatomic medullary locking (AML) total hip arthroplas
ty, using the acetabular cup system (ACS). The mode of failure was by a com
bination of wear-through to the metal backing and liner fracture. Four of t
he 5 cases were associated with a 32-mm head. The ACS design is flawed beca
use of a lack of hemispherical geometry resulting in rim loading on a regio
n of the liner that has the thinnest polyethylene. All patients with this t
ype of implant should be monitored for evidence of excessive polyethylene w
ear.