Differential non-destructive image current detection in a Fourier transform quadrupole ion trap

Citation
Er. Badman et al., Differential non-destructive image current detection in a Fourier transform quadrupole ion trap, J MASS SPEC, 34(8), 1999, pp. 889-894
Citations number
32
Categorie Soggetti
Chemistry & Analysis","Spectroscopy /Instrumentation/Analytical Sciences
Journal title
JOURNAL OF MASS SPECTROMETRY
ISSN journal
10765174 → ACNP
Volume
34
Issue
8
Year of publication
1999
Pages
889 - 894
Database
ISI
SICI code
1076-5174(199908)34:8<889:DNICDI>2.0.ZU;2-F
Abstract
Dual-detector differential non-destructive Fourier transform detection in a quadrupole ion trap is shown to improve signal intensity and reduce noise compared with spectra recorded using a single detector. A larger area detec tor in each end-cap electrode is machined to fit its hyperbolic shape and s o minimize field imperfections on the z-axis, Argon, acetophenone and bromo benzene spectra were recorded to allow a comparison between single- and dua l-detector (differential) modes of detection and to demonstrate the improve ment achieved with differential detection, Copyright (C) 1999 John Wiley & Sons, Ltd.