P. Karaiskos et al., A Monte Carlo investigation of the dosimetric characteristics of the VariSource Ir-192 high dose rate brachytherapy source, MED PHYS, 26(8), 1999, pp. 1498-1502
Citations number
13
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
An analytical Monte Carlo simulation code, incorporating in detail source c
onstruction and dimensions, was used to investigate the dosimetric characte
ristics of the VariSource Ir-192 high dose rate brachytherapy source. Dose-
rate constant, radial dose functions, and anisotropy functions, utilized in
the AAPM Task Group 43 dose estimation formalism, were calculated with the
source centered in a spherical water phantom of 30 cm in diameter. The res
ults, which are in agreement with the corresponding data available in the l
iterature, are compared to those obtained in our previous study for the wid
ely used microSelectron Ir-192 high dose-rate brachytherapy source. The dos
e-rate constant was found to be equal to 1.043+/-0.005 cGy h(-1) U-1 for th
e VariSource, compared to a value of 1.116+/-0.006 cGy h(-1) U-1 calculated
for the microSelectron. Significant differences in the anisotropy of the t
wo sources are observed only for polar angles close to their long axis and
are due to their different dimensions. (C) 1999 American Association of Phy
sicists in Medicine. [S0094-2405(99)03008-4].