Disopyramide ion-selective poly(vinyl chloride) membrane electrodes contain
an ion-pair complex of disopyramide tetrakis (4-chlorophenyl)borate as the
electroactive material with 2-nitrophenyl octyl ether (electrode I) or bis
(2-ethylhexyl)sebacate (electrode II) as the solvent mediators. The electro
des (I and II) show a linear response for the disopyramide concentration ra
nge 3.0 x 10(-4)-10(-2) and 8.5 x 10(-5)-10(-2) mol.l(-1) over a pH range o
f 3.75-8.30 with a cationic slope of 57.3 and 58.5 mV decade(-1), respectiv
ely, The response time varied from 20 s to 1 min depending on the disopyram
ide concentration. Electrode II was used for the potentiometric determinati
on of the disopyramide phosphate substance and the content of capsules with
average recovery and mean standard deviation (+/- SD) of 100.8 +/- 0.34 an
d 100.1 +/- 0.68 of the nominal values, which are comparable with those obt
ained by the U.S. Pharmacopoeia method.