An improved technique for comparing Gamma Knife dose-volume distributions in stereotactic radiosurgery

Citation
Sm. Tozer-loft et al., An improved technique for comparing Gamma Knife dose-volume distributions in stereotactic radiosurgery, PHYS MED BI, 44(8), 1999, pp. 1905-1919
Citations number
21
Categorie Soggetti
Multidisciplinary
Journal title
PHYSICS IN MEDICINE AND BIOLOGY
ISSN journal
00319155 → ACNP
Volume
44
Issue
8
Year of publication
1999
Pages
1905 - 1919
Database
ISI
SICI code
0031-9155(199908)44:8<1905:AITFCG>2.0.ZU;2-6
Abstract
A function derived from the geometry of brachytherapy dose distributions is applied to stereotactic radiosurgery and an algorithm for the production o f a novel dose-volume histogram, the Anderson inverse-square shifted dose-v olume histogram (DVH), is proposed. The expected form of the function to be plotted is checked by calculating its value for single focus exposures, an d its application to clinical examples of Gamma Knife treatments described. The technique is shown to provide a valuable tool for assessing the adequa cy of radiosurgical plans and comparing and reporting dose distributions.