The application of factor analysis to solvatochromic parameters and pH(s) values for the standardization of potentiometric sensors in mobile phases used in liquid chromatography

Citation
J. Barbosa et al., The application of factor analysis to solvatochromic parameters and pH(s) values for the standardization of potentiometric sensors in mobile phases used in liquid chromatography, TRAC-TREND, 18(8), 1999, pp. 543-549
Citations number
34
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
TRAC-TRENDS IN ANALYTICAL CHEMISTRY
ISSN journal
01659936 → ACNP
Volume
18
Issue
8
Year of publication
1999
Pages
543 - 549
Database
ISI
SICI code
0165-9936(199908)18:8<543:TAOFAT>2.0.ZU;2-8
Abstract
The pH(s) values for reference buffer solutions in aqueous acetonitrile, me thanol, and tetrahydrofuran mixtures used as mobile phases in LC were subje cted to factor analysis in order to discover the number of factors which af fect the variation with the percentage of organic co-solvent of the whole d ata sets. Afterwards, target factor analysis was used to identify these fac tors. The pH(s) values were correlated with the solvatochromic parameters ( pi*, alpha, beta, and E-T) Of the hydro-organic mixtures according to the r esults obtained using these chemometric techniques. An equation is proposed which permits the calculation of pH(s) values in any aqueous acetonitrile, methanol, or tetrahydrofuran mixtures, and thus accurate pH measurements c an be made in these media as well as in water. An overview of the effects o n pH(s) values of preferential solvation in these media is also presented. (C) 1999 Elsevier Science B.V. All rights reserved.