J. Upadhyay et al., A simple power supply and control unit for pulsed operation of a microchannel plate imaging detector, SADHANA, 24, 1999, pp. 235-243
Citations number
2
Categorie Soggetti
Engineering Management /General
Journal title
SADHANA-ACADEMY PROCEEDINGS IN ENGINEERING SCIENCES
A simple power supply and control electronic system to operate microchannel
plate (MCP) detectors for imaging transient events such as X-ray emission
from laser-produced plasmas is described. The power supply consists of two
parts, viz. a high voltage pulse generator of 1 kV amplitude with a variabl
e pulse duration of 100 mu s to 1 ms for the MCP input, and a regulated 5 k
V d.c. for biasing the phosphor screen. The control unit synchronizes the h
igh voltage pulse with the event and operates a safety switch to guard the
detector against any accidental rise in the background pressure. A signal-t
o-noise ratio greater than or equal to 75 is obtained, limited by the noise
contribution of the Charge Coupled Device (CCD)-frame grabber combination
attached to the MCP. The system is compact, has provision for both positive
and negative pulses, is extendable to higher voltages for operating two-st
age MCP, and is immune to electromagnetic interference.