Jd. Wang et al., THE INFLUENCE OF THE TEMPERATURE OF BLACKBODY FOR CALIBRATING FTIR SYSTEM ON THE INSTRUMENT RESPONSE FUNCTION, Spectroscopy letters, 30(4), 1997, pp. 783-791
The instrument response function of Fourier transform infrared spectro
meter is a very important parameter in infrared emission spectroscopy
to obtain infrared spectral radiance distribution of the desired sampl
e, the concentrations of the components of combustion products in the
flame etc. This paper gives the effect of the temperature of the stand
ard blackbody for calibrating Fourier transform infrared spectrometer
on the instrument response function. The result illustrates that the v
alues of the instrument response functions decrease as the blackbody t
emperature increased, and tends to a limit. The paper points out that
the condition must be observed for using instrument response function
correctly.