Ms. Lourens et al., Flow-volume curves as measurement of respiratory mechanics during ventilatory support: the effect of the exhalation valve, INTEN CAR M, 25(8), 1999, pp. 799-804
Objective: To assess the feasibility of expiratory flow-volume curves as a
measurement of respiratory mechanics during ventilatory support: to what ex
tent is the shape of the curve affected by the exhalation valve of the vent
ilator?
Design: Prospective, comparative study.
Setting: Medical intensive care unit of a university hospital.
Patients: 28 consecutive patients with various conditions, mechanically ven
tilated with both the Siemens Servo 900C and 300 ventilators, were studied
under sedation and paralysis.
Interventions: The ventilator circuit was intermittently disconnected from
the ventilator at end-inspiration in order to obtain flow-volume curves wit
h and without the exhalation valve in place.
Measurements and results: Peak flow (PEF) and the slope of the flow-volume
curve during the last 50% of expired volume (SF50) were obtained both with
and without the exhalation valve in place. The exhalation valve caused a si
gnificant reduction in peak flow of 0.3 l/s (from 1.27 to 0.97 l/s) with th
e Siemens Servo 900 C ventilator and of 0.42 l/s (from 1.36 to 0.94 l/s) wi
th the Siemens Servo 300 ventilator (p < 0.001). The SF50 was not affected.
Conclusion: In mechanically ventilated patients, the exhalation valve cause
s a significant reduction in peak flow, but does not affect the SF50. This
study further suggests that the second part of the expiratory flow-volume c
urve can be used to estimate patients' respiratory mechanics during ventila
tory support.