Sar. Kannengiesser et al., Geometric distortion correction in gradient-echo imaging by use of dynamictime warping, MAGN RES M, 42(3), 1999, pp. 585-590
Citations number
12
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
Distortion caused by local static field Variations in magnetic resonance im
aging can be corrected by using a method developed by Chang and Fitzpatrick
(CF) that requires two acquisitions identical except for the polarity of t
he read gradient, This approach has previously been applied to spin-echo im
aging, In this report, its role in correcting geometric distortion in gradi
ent-echo imaging is investigated using a dynamic programming algorithm, Thi
s approach appears to successfully handle many of the edge artifacts otherw
ise inherent in the CF method, Once the correction has been made, it is the
n possible to extract the local magnetic field itself, (C) 1999 Wiley-Liss,
Inc.