Hl. Anderson et al., INFLUENCE OF TIME CONSTANTS OF SENSORS ON EVALUATION OF THERMAL-ANALYSIS DATA, Journal of thermal analysis, 48(4), 1997, pp. 769-782
All temperature sensors have a finitely time constant. The influence o
f the sensor time constant tau(s) on the results of kinetic evaluation
is demonstrated at four reaction types. The ignorance of the sensor i
ndolence gives incorrect activation parameters. Therefore the determin
ation of tau(s) is necessary. For the estimation of parameters the non
linear evaluation program TA-kin was used. With its help it is possibl
e to find the real parameters, also when tau(s)=32 s, if the real tau(
s)-value was entered.