A. Amirav et S. Dagan, A DIRECT SAMPLE INTRODUCTION DEVICE FOR MASS-SPECTROMETRY STUDIES ANDGAS-CHROMATOGRAPHY MASS-SPECTROMETRY ANALYSES, European mass spectrometry, 3(2), 1997, pp. 105-111
Citations number
11
Categorie Soggetti
Spectroscopy,"Physics, Atomic, Molecular & Chemical
A novel direct sample introduction (DSI) device is described, based on
sample introduction in a disposable mini-vial (test tube) inside the
existing gas chromatography (GC) injector, This DSI device effectively
transforms a conventional GC injector, followed by a short capillary
column, into a very cost-effective alternative to the standard direct
insertion probe and enables a continuous sample flux at the mass spect
rometer ion source for mass spectrometry (MS) and MS-MS studies, This
same device also enables direct introduction of dirty samples for GC a
nd GC-MS analysis, The dirty sample introduction is based on the therm
al extraction of semi-volatile sample compounds inside the GC injector
, while the non-volatile residue is retained in the sample vial that i
s disposed of after the analysis, The major capabilities of this new D
SI device are demonstrated and its various features are discussed.