Repeatability and reproducibility of retention data and band profiles on reversed-phase liquid chromatography columns III. Results obtained with Kromasil C-18 columns
M. Kele et G. Guiochon, Repeatability and reproducibility of retention data and band profiles on reversed-phase liquid chromatography columns III. Results obtained with Kromasil C-18 columns, J CHROMAT A, 855(2), 1999, pp. 423-453
The reproducibility of the retention data and the band profiles was investi
gated with Kromasil C-18 columns (silica-based monomeric type reversed-phas
e packing material). High precision data were obtained and statistically co
mpared among five columns from the same batch (column-to-column reproducibi
lity) and six columns, one from each of six different batches (batch-to-bat
ch reproducibility). These data were acquired under five different sets of
chromatographic conditions, for a group of 30 neutral, acidic and basic com
pounds selected as probes following an experimental protocol previously des
cribed. Data characterizing the retention time, the retention factor, the s
eparation factor, the column efficiency and the peak asymmetry for the diff
erent probe compounds are reported. Factors describing the silica surface i
nteraction with the selected probe compounds, such as the hydrophobic inter
action selectivity, the steric selectivity, and the separation factors of b
asic compounds at different pH values were also determined. The influence o
f the underlying silica on these data and correlations between the chromato
graphic and physico-chemical properties of the different batches are discus
sed. (C) 1999 Elsevier Science B.V. All rights reserved.