A flow-through solid phase spectrophotometric (SPS) sensing device is propo
sed for the determination of minoxidil. The analyte is concentrated on Seph
adex SP-C25 ion-exchanger packed in a flow cell and it is monitored by UV-V
is spectrophotometry at 282 nm, without derivatization reaction. When a HCl
(10(-2) mol l(-1))/NaCl (5 x 10(-2) mol l(-1)) solution is used as carrier/
desorbing solution, the sensor responds linearly in the measuring range of
0.2-7, 0.1-4 and 0.05-2 mu g ml(-1) with detection limits of 60, 33 and 6 n
g ml(-1) for 600, 1000 and 2000 mu l of sample, respectively. The relative
standard deviations (%) for these volumes are 0.38, 1.06 and 2.63, respecti
vely. The method was satisfactorily applied to the determination of minoxid
il in pharmaceutical preparations and the results were compared with those
obtained by high performance liquid chromatography (HPLC). (C) 1999 Elsevie
r Science B.V. All rights reserved.