M. Derbyshire et al., Optimization of the simultaneous determination of Cr(VI) and Cr(VI) by ionchromatography with chemiluminescence detection, ANALYT CHEM, 71(19), 1999, pp. 4203-4207
An optimized method for the simultaneous determination of Cr(III) and Cr(VI
) in aqueous solutions using ion chromatography with chemiluminescence dete
ction is described. Excellent resolution of the two chromium species was ob
tained using a single mixed-bed ion-exchange column with continuous elution
. After postcolumn reduction of Cr(VI) to Cr(III), the light emitted during
the Cr(III)-catalyzed oxidation of luminol with hydrogen peroxide was meas
ured. Parameters affecting the postcolumn reactions such as reductant conce
ntration, reductant mixing, point of luminol introduction, and luminol flow
rate were optimized. The calibration curves in the range tested (0.01-50 m
u g L-1) were linear, and detection limits of 0.002 pg L-1 for both Cr(III)
and Cr(VI) were obtained. The results of the analyses of the water referen
ce materials LGC CRM6010 and NIST SRM1643d with certified chromium values o
f 49 +/- 4 and 18.53 +/- 0.20 Pg L-1 and found to contain only Cr(III) were
49.2 +/- 1.8 and 19.0 +/- 1.5 mu g L-1, respectively. Values of 10.6 +/- 0
.5 and 10.1 +/- 0.5 mu g L-1 were obtained when a simulated water sample co
ntaining 10 mu g L-1 Cr(III) and Cr(VI) was analyzed.