Je. Prest et Pr. Fielden, The simultaneous separation of twenty-one metal cations using capillary isotachophoresis with on-column conductivity detection, ANAL COMMUN, 36(9-10), 1999, pp. 333-335
The possibility of using capillary isotachophoresis coupled with on-column
conductivity detection for the separation of complex metal cation mixtures
has been demonstrated. The simultaneous separation of a mixture of the 21 m
etal cations Ca2+, Mg2+, Mn2+, Co2+, Ni2+, Zn2+, Be2+, La3+, Ce3+, Pr3+, Nd
3+, Sm3+, Eu3+, Gd3+, Cu2+, Y3+, Ho3+, Er3+, Tm3+, Yb3+ and Lu3+ has been r
eproducibly achieved using aqueous electrolytes. Calibration curves were pr
oduced for two of these ions, Co2+ and Er3+. A linear response was observed
over the range from 0.02 mmol dm(-3) to 1.70 mmol dm(-3) for cobalt and 0.
05 mmol dm(-3) to 1.00 mmol dm(-3) for erbium. These curves allowed limits
of detection to be calculated as 0.015 mmol dm(-3) for cobalt and 0.045 mmo
l dm(-3) for erbium.