M. Hirose et al., Usefulness of newly designed pigtail catheter with multiple side holes by reducing incidence of ventricular ectopy, CATHET C IN, 48(2), 1999, pp. 220-225
This study was designed to determine whether a newly designed 5 Fr pigtail
catheter of Jet Balance with multiple side holes reduces the incidence of v
entricular ectopy during left ventriculography, To examine the catheter mov
ement in vitro, we injected a colored glycerin solution by an injector thro
ugh 5 Fr Jet Balance catheter and 6 Fr conventional pigtail catheter suspen
ded in a water filled, glass container, A recoil of the catheter caused by
injection was observed for a 6 Fr conventional catheter; however, no cathet
er movement was observed when a 5 Fr Jet Balance was used. To evaluate clin
ical usefulness of the 5 Fr Jet Balance, we compared the number of incidenc
e of ventricular ectopy during left ventriculography between 5 Fr Jet Balan
ce and 6 Fr conventional pigtail catheters, In 104 consecutive patients who
underwent left ventriculography, a 5 Fr Jet Balance was used in 55 patient
s (5 Fr Jet Balance group), and a 6 Fr conventional pigtail in 49 patients
(6 Fr conventional pigtail group). The 5 Fr Jet Balance group had a signifi
cantly lower incidence of VPB (21.8% vs. 42.9%, P = 0.02), ventricular tach
ycardia (1.9% vs. 14.3%, P = 0.02), and multiple VPB (9.0% vs. 34.7%, P = 0
.001), respectively, than the 6 Fr conventional pigtail group. In conclusio
n, a 5 Fr Jet Balance pigtail catheter is useful and safe as regard to the
incidence of ventricular ectopy during left ventriculography. (C) 1999 Wile
y-Liss, Inc.