We measured the optical and the X-ray transmission of the optical blocking
filters for the X-ray Imaging Spectrometers (XISs) which are the X-ray CCD
cameras of the ASTRO-E satellite. We conclude that the oxidation of the alu
minum reduces the optical transmission down to similar to 60-70% of the the
oretical value of the aluminum. We achieved optical transmission below 5 x
10(-5) in the range from 4000 to 9500 Angstrom by using aluminum thickness
of 1200 Angstrom, while the theoretical calculation requires 800 Angstrom.
The measurement of absolute quantum efficiency of XIS is also performed at
several particular energies. We confirmed 20% quantum efficiency at 0.5 keV
for the XIS engineering model (XIS EM). (C) 1999 Elsevier Science B.V. All
rights reserved.